Gene Pulser Xcell™ Electroporation Systems

The Gene Pulser Xcell™ system is a modular electroporation system for transfecting every cell type. The system includes a main unit, a ShockPod™ cuvette chamber, and your choice of accessory modules: the capacitance extender (CE module) and the pulse controller (PC module)

Features and Benefits

·         Universal electroporation — transfects all cell types, from primary and stem cells to bacteria and yeast

·         Preset protocols — include the most common mammalian and bacterial cell types

·         Flexibility — choice of programs for preset protocols, optimization protocols, manual operation, and/or user protocols

·         Protocol library — collection of electroporation protocols for every cell type including primary, immortal, and bacterial cells

·         Data management — enables storage and recall of parameters used in the previous 100 experiments, useful for troubleshooting

·         Reproducibility — uses PulseTrac™ circuitry and arc-protection features to ensure reproducibility and sample protection

Gene Pulser Xcell™ Total System
Includes main unit, CE module, PC module, and ShockPod™ cuvette chamber
 
Outputs Waveform: Exponential or square
Voltage: 10–3,000 V
Capacitance 10–500 V, 25–3,275 µF in 25 µF increments
500–3,000 V, 10, 25, 50 µF
Resistance (parallel) 50–1,000 Ω in 50 Ω increments, plus infinity
Sample resistance 20 Ω minimum at 10–2,500 V
600 Ω minimum at 2,500–3,000 V
Square-wave timing  10–500 V: 0.05–10 ms in 0.05 ms increments, 10–100 ms pulse in 1 ms increments, 1–10 pulses, 0.1–10 sec interval
500–3,000 V: 0.05–5 ms in 0.05 ms increments, 1–2 pulses, 5 sec minimum interval
Gene Pulser Xcell Eukaryotic System
Includes main unit, CE module, and ShockPod cuvette chamber
 
Outputs and other specifications Same as total system without the parallel resistance
Gene Pulser Xcell Microbial System
Includes main unit, PC module, and ShockPod cuvette chamber
 
Outputs Waveform: Exponential or square
Voltage: 200–3,000 V
Capacitance  10, 25, 50 µF
Resistance (parallel)  50–1,000 Ω in 50 Ω increments, plus infinity
Sample resistance  20 Ω minimum at 200–2,500 V
600 Ω minimum at 2,500–3,000 V
Square-wave timing  0.05–5 ms pulse duration in 0.05 ms increments,
1–2 pulses, 5 sec minimum interval
Gene Pulser Xcell Main Unit  
Outputs Waveform: Exponential or square
Voltage: 200–3,000 V
Capacitance 10, 25, 50 µF
Sample resistance 20 Ω minimum at 200–2,500 V
600 Ω minimum at 2,500–3,000 V
Square-wave timing  0.05–5 ms pulse duration in 0.05 ms increments,
1–2 pulses, 5 sec minimum interval
General  
Input voltage 100–120 VAC or 220–240 VAC, 50/60 Hz
Power Maximum 240 W (during short charging periods)
Operating environment Temperature 0–35°C, humidity 0–95% (noncondensing)
Regulatory Safety EN 61010, EMC EN 61326 Class A
Dimensions (W x D x H) Main unit: 31 x 30 x 14 cm
CE module: 31 x 28 x 9 cm
PC module: 31 x 28 x 5 cm
Weight Main unit: 6.6 kg
CE module: 3.1 kg